Quick Start 7
FastCalXP Operation Manual
TAKE THE AS RECEIVED DATA
Perform a New Calibration.
1 Click the Perform As Received Cal button.
2 The Calibration Wizard will auto-detect the reference as described in the discussion of References.
If a new r
eference is detected, the Calibration Wizard will prompt you to enter the current certi cate information for the reference. If using an nVision with
two modules you will be prompted to select one of the modules to use for the calibration.
3 If the pro le requires it, exercise the gauges by increasing and decreasing the pressure over the full range of the gauge one or more times
. Refer to the
GaugeCalHP Operation Manual or the discussion of Exercising on page 17 for more information.
The C
alibration Window will then open. Refer to description of the Calibration Window in the Menus and Windows section.
4 Select Actual or Peak mode and step through the calibration. Refer to the discussions of Calibration Strategies, Actual Mode Calibration, and Peak Mode
C
alibr
ation later in the manual.
Load Existing As Received Data.
If the As Received calibration has been completed in a previous session, complete the As Received Step by clicking the Load Previous Data button.
TAKE THE AS LEFT DATA
If the gauge failed the As Received calibration and adjustments have been made to the gauge, click on the Perform As Left Cal button. Follow steps 2 through
4 in the Perform a New Calibration instructions above.
Alternativ
ely, if the gauge passed the As Received calibration and no adjustments have been made to the gauge, FastCalXP can use the As Received calibra-
tion for the As Left calibration. Click on the Use As Received for As Left button.
GENERATE AND PRINT THE CERTIFICATE
1 Enter any notes regarding the calibration or modi cations made to the gauge in the Calibration Notes /Modi cations box and click on the Save Data button
in the Calibration Status box. Notes cannot be added after the certi cat
e has been generated.
2 Create the calibration certi cat
e by clicking on the Generate Certi cate button. FastCalXP will create the certi cate and open the certi cate by running
Excel. Go to the Excel window with the current certi cate and print the certi cate.
Refer to the Creating Certi cate
s section for more information on working with certi cates.