Calibrating Gauges 11
FastCalXP Operation Manual
8 Select the Accuracy of the gauge from the pull-down list or specify a % of span or Custom accuracy.
Refer to the discussion of Models on page 36 in the Managing the Database section.
9 Click Next to select the As Received and As Left calibration Pro les.
If the desired pro les are not listed, click on the Create New Pro le button t
o open the Pro les editor. Refer to the discussion of Test Pro les on page 39.
10 Select a location from the pull-down list or click on the Add New Location button.
A location is not required. However, it is commonly used to specify the operating location of the gauge within the user’s facilit
y or
, by a calibration labora-
tory, for customer information. Use the boxes provided to enter speci c location information and notes