Module Queries and Commands 5
nVision Programming Instructions
Module Queries and Commands 5
X
Message Store
MOD:MSG? #
X
Model
MOD:MOD? #
X
Reading
MOD:RD? #
Note: The MA20 switch test states are “1” for open, and “0” for closed. Units for MA20 module are either “mA, % 4-20mA, %10-50mA
(for MA-20+ modules only), VDC, or Switch Test.
X
Serial Number
MOD:SN? #
X
Unit
MOD:UNIT? #
Response: PSI, kg/cm2, inHg, inH2O, mmHg, mmH2O, kPa, bar, mbar, MPa, user, C, F, R, K, Ohm, mA, %4-20mA, %10-50mA, VDC, or SwitchTest
Note: This query is only available when the nVision has R080007 or greater rmware.
X
Water Reference
MOD:H2O? #
Response: 4C, 60F, or 68F
Note: This query is only available when the nVision has R080007 or greater rmware and is populated with a pressure module (PM).